IMPSPEC

IMPSPEC is a series of tools for visualizing and analyzing survey spectroscopy data to obtain line brightness time histories.  The goal is not to extract all of the information from the spectra, but provide an easy interface to extract the most relevant bits to inform plasma operations and physic studies that are interested in tracking impurity contamination.  Data are stored in an intuitive way in MDSPlus with descriptors of the emission, allowing for easy integration into DWSCOPE and JSCOPE.

The code is written in IDL and kept in /p/nstxusr/nstx-users/mreinke/GENIE/IMPSPEC/ under version control from the central GITHUB repository.  This directory is readable from skylark, allowing automated between shot analysis and manual creation of tree nodes.

IMSPEC Compatible Diagnostcs


IMPSPEC was written to analyze data from the XEUS and LoWEUS spectrometers as deployed at Alcator C-Mod.  These and other similar flat-field VUV/SXR survey spectroscopy have limited spectral resolution so detailed analysis of Doppler or Zeeman features is not possible.  This makes for a simplified analysis where the main goal is to extract a time evolving brightness of a single line from a region of contaminating neighboring lines. 

NSTX-U compatible diagnostics would be:  XEUS, LoWEUS, MONA-LISA and SPRED.

Presently there is no ability to deal with multi-channel spectroscopy such as from imaging VUV system (TGIS), but this is expected to be integrated at some point in the future.  Similarly, an interest to expand to deal with visible spectroscopy is anticipated.

.ispec file format



A base set of ASCII files are located in  /IMPSPEC/ispec/ with the element name and suffix .ispec.  An example of N.ispec is below:

N,7
nlines=3

###NOTES###
line=0                          ;line indicator
220.0,222.0                     ;wavelength range to fit
inst=0.2                        ;instrumental width
nlines=1                        ;number of lines to fit
221.15,Ar XV,Be                 ;primary line of interest then interfering lines
###DATA###

line=0
spec=xeus
24.5,25.2
inst=0.05
nlines=5
24.782,N VII,H
24.9006,N VI (n=3),He
24.8640,Ar XVI,Li
25.0240,Ar XVI,Li
24.7400,Ar XV,Be

line=1
spec=xeus
28.5,29.8
inst=0.05
nlines=2
28.7861,N VI,He
29.0835,N VI,He

line=2
spec=xeus
20.6,21.25
inst=0.05
nlines=1
20.9100,N VII (n=3),H
###END###


There are three sections:  header, notes and data

HEADER
The header is very simple, containing the element abbreviation and atomic number on the first line and the number of lines that the ispec file contains

N,7
nlines=3

NOTES
The notes section is completely free-form.  Any information can be included and is meant for the user to offer any description.  By default a description of the line information is included as a remindar.  Nothing in this section impacts the operation of IMPSPEC - so feel free to use it however you see fit!  This section is ended by ###DATA###

###NOTES###
line=0                          ;line indicator
220.0,222.0                     ;wavelength range to fit
inst=0.2                        ;instrumental width
nlines=1                        ;number of lines to fit
221.15,Ar XV,Be                 ;primary line of interest then interfering lines

DATA
This section contains the data that will be used to fit.  Multiple instruments can be called, the example here only looks at the primary He-like (N VI) and H-like (N VII) nitrogen transitions.  There are text blocks for each of the nlines  specified in the header.  Each block contains a variable length amount of data depending on the complexity of the spectra to be analyzed.  How these data are used is described in "How IMPSPEC Works".  This section is ended by ###END###.
###DATA###

line=0
spec=xeus
24.5,25.2
inst=0.05
nlines=5
24.782,N VII,H
24.9006,N VI (n=3),He
24.8640,Ar XVI,Li
25.0240,Ar XVI,Li
24.7400,Ar XV,Be

You can generate alternative .ispec files for non-standard shots where you might want to fit spectral features not normally seen.  For example, compare above to /p/nstxusr/nstx-users/mreinke/GENIE/IMPSPEC/ispec/N.ispec.long which includes a number of N IV and N V lines on LoWEUS.
 

How IMPSPEC Works


The data for each line are used to isolate a portion of specific diagnostic's spectrum and fit a constrained multi-peak Gaussian fitting to the data.
  • spec identifies which spectrometer to use (these names are hard-coded into IMPSPEC.PRO)
  • XX.X, YY.Y sets the wavelength range to be used from the spectrometer
  • inst sets the maximum instrument width
  • nlines sets the number of individual lines to be included in the multi-peak fit
  • next are descriptions of the individual lines that have the the following comma seperated data
    • center wavelength of line
    • label to identify the line
    • isoelectronic state of the line
The IDL code MPFIT is used to do the multi-peak fit with the following constraints:
  1. all lines have the same width, with a maximum set by inst
  2. all lines have the the same relative spacing, thus move together
  3. all lines must have a positive peak height and all are independent.

IMPSPEC Tree


example of IMPSPEC tree layout

In the PASSIVESPEC tree, there is an IMPSPEC node, below which there are an arbitrary number of element nodes (ie B, N, etc), and a PROF node.

In each of the elements nodes, there are the following:

 NODESTATUS
 I/ODESCRIPTION
 CONVunused
 input    time-evolving convection for impurity transport modeling
 CSDENunused
 output    result of time-evolving impurity transport simulation
 DIFFunused
 input    time-evolving diffusion for impurity transport modeling
 ISPECin-use    
input
    local, shot-specific copy of the ISPEC file
 NLINESin-use
input
    number of lines
 RUNISPECin-use    
input
    action node to automate IMPSPEC on shot cycle
 LINE#in-use
output
    sub-node containing analysis for each 
 PROF   
 unused 
 input           profile data to drive the impurity transport simulation

Each of the LINE nodes contains the following:

 NODESTATUS
I/O
DESCRIPTION
 BR in-use output    line brightness(es) from running IMPSPEC
 BR_MOD unused output    modeled line brightness from the transport simulation
 COEFS in-use output    gaussian fit coefficients from IMPSPEC
 DLAM in-use input    wavelength region fit
 ETA unused input    'calibration'  constant to convert counts/s to ph/s/m^2
 LABEL in-use input    text label(s) for the line(s) included in the fit
 LAM in-use input    center wavelength(s) for the line(s) included in the fit
 NZ unused output    using BR, BR_MOD and eta, determine the time-evolving impurity     density from the time-evolving impurity transport simulations



As can be seen, there are a number of features laid out in the tree structure that have yet to be implemented.  The goal is to have the capability for between shot impurity transport simulation to estimate the brightness time history per unit impurity density.  This would then allow the impurity density to be estimated from the line brightness, correcting for temperature and density dependence in the brightness.  Presently only the capability to compute the line brightness is implemented.

Creating or Modifying IMPSPEC trees



To load GENIE (and IMPSPEC) on skylark so that tree modifications can be made:
@/home/mreinke/skylark_genie_startup.pro

ADD_IMPSPEC,shot,z=z is used to add the PASSIVESPEC.IMPSPEC node and the corresponding element nodes for the 1D INTARR, z.  For example, ADD_IMPSPEC,shot,z=[7,8] would add the IMPSPEC.N and IMPSPEC.O nodes and fill them with the relevant number of lines in the .ispec file.  This same program can be used to remove it via ADD_IMPSPEC,shot,/rm

These will use the /p/nstxusr/nstx-users/mreinke/GENIE/IMPSPEC/ispec/X.ispec files by default.  You can use the WRITE_ISPEC2NODE,shot,path,/rm,/tree to overwrite the IMPSPEC.X node with the structure outlined in the X.ispec file defined by the path.  The /tree keyword automatically puts the new X.ispec into the tree.  This process allows new X.ispec stored in the users local directory can be added.

If you do not need to rewrite the tree structure, i.e. if you're just tweaking settings within the same number of lines, then the X.ispec file can be written to the tree using WRITE_ISPEC2TREE,shot,path. 

Running IMPSPEC Manually and Viewing Results


To load GENIE (and IMPSPEC) on nstxpool run @/p/nstxusr/nstx-users/mreinke/GENIE/mach/nstxu/genie_ini.bat in IDL.

Spectral analysis is done for each element separately, using RUN_IMSPEC,shot,z.  In this way, the entire set of impurities can be run in-parallel, allowing all the line brightness data to be analyzed between shots.  Data are written to the tree.

Scope

A DWSCOPE for viewing the results is in /u/mreinke/scopes/impspec.dat 

This needs to be manually updated as the number of elements is expanded

Widget

A widget to view the line fits for each Z and LINE# is in /p/nstxusr/nstx-users/mreinke/GENIE/widgets/w_impspec.pro  This can be compiled after running the above genie_ini.bat and then run using w_impspec.  This will also need some manual updates as the number of lines are updated, or the zlist optional input can be used to set what impurities to look for.

example of impspec widget

***need to input details on how to use this widget***  2/10/16

Running IMPSPEC on the NSTX-U Shot Cycle


TBD once the LLNL suite is installed!  Currently IMPSPEC is not in the model tree and will be added once the spectrometers are installed.  There are shell scripts that will be run for each element.

Comments


IMPSPEC (and this page) is maintained by M.L. Reinke (mreinke@pppl.gov) x3908.  Please send any comments or questions and bug reports to me or stop by B236!